EuroPCR2026-images-104955-Whats

What complication has occurred?
The wire is subintimal
Fractured microcatheter
Retained balloon tip
Correct! The correct answer is "Retained balloon tip".
 
Fluoroscopy showed a free floating balloon tip in the proximal PL branch as the balloon tip had detached from the shaft. This was trapped out by railroading a Guide extension and using a 3.0 balloon.

 
Good try! The correct answer is "Retained balloon tip".
 
Fluoroscopy showed a free floating balloon tip in the proximal PL branch as the balloon tip had detached from the shaft. This was trapped out by railroading a Guide extension and using a 3.0 balloon.
 
 
{"name":"EuroPCR2026-images-104955-Whats", "url":"https://www.poll-maker.com/QYXHCLKX1","txt":"What complication has occurred?, Correct! The correct answer is \"Retained balloon tip\".   Fluoroscopy showed a free floating balloon tip in the proximal PL branch as the balloon tip had detached from the shaft. This was trapped out by railroading a Guide extension and using a 3.0 balloon., Good try! The correct answer is \"Retained balloon tip\".   Fluoroscopy showed a free floating balloon tip in the proximal PL branch as the balloon tip had detached from the shaft. This was trapped out by railroading a Guide extension and using a 3.0 balloon.","img":"https://www.poll-maker.com/3012/images/ogquiz.png"}